LABORATORY 7
Unknowns No. 2 and No. 3
Objectives: During this laboratory, you will:
Sequence of Student Work and Observations
B. Unknown No. 3
Student work: Day 2 of Unknown No. 3
Read the instructions carefully before inoculating the kits.
Record the inoculation time. The tests must be read within 18-20 hours. You will be responsible for reading and recording your results in that time frame. Your instructor will tell you where to meet in order to read the results.
The Crystal Enteric Identification System is a miniaturized system of tests for identification of enteric organisms. These tests are modifications of classical methods that measure metabolic endproducts and enzyme activities. These tests include tests for fermentation, oxidation, degradation, and hydrolysis of various substrates. There are also chromogen-linked substrates to detect enzymes that microbes use to metabolize various substrates.
After inoculation with an isolated colony from a clinical specimen (you will use an isolated colony from your Unknown No. 3), the system is incubated for 18-20 hours at 37o C. The system is then examined for color changes for each substrate included. Color changes result from metabolic activities of the microbe. The resulting pattern of the 30 reactions in the system is converted into a ten digit profile number that is used as the basis for identification using a code book of profiles. Proceed as follows using Unknown No. 3:
1. Remove Crystal lids from pouch, and discard the dust cover and desiccant.

2. Place one (no more) loop-full of Unknown No. 3 into a tube of Enteric/Stool ID Inoculum Fluid; then recap tube and mix well for 15-20 seconds to evenly suspend bacterial cells.
3. Mark Crystal Base with your name, date, and the code of your Unknown No. 3.
4. Pour the entire content of inoculum into the target area (reservoir)
of the base.

5. Distribute the inoculum by holding the base in both hands; tilt the base to roll the inoculum gently along the tracks until all of the wells are filled. Roll back any excess fluid to the target area and place the base on the bench. Place cotton over target to absorb fluid.

6. Align the Crystal lid so that the labeled end of the lid is
on top of the target area of the base.

7. Push the lid down until a slight resistance is felt; then, place
thumbs on edges of the lid towards the middle of the panel and push downwards
until the lid snaps into place (Listen—2 clicks).

8. Place your Crystal Enteric Identification System in the container
as per instructor's direction.